Overview
FEATURE
- Temperature ranges 30℃ to 50℃
- Target size is 72mm x 72mm
- Emissivity is 0.97 ± 0.02
- Accuracy is better than 0.1℃
- Control stability is 0.1℃ after 10 minutes
- Heat up time is 2-10 minutes depends
- Built-in temperature correction function
APPLICATION
- Ambient temperature compensation for thermal image camera, assist thermal image camere test the body temperautre more much accurate in public area.
The BL670 blackbody calibrator is a small size blackbody calibration source utilizing a digital indicating temperature ambient temperature + 5℃ to 50℃. A precision platinum RTD temperature sensor is embedded in the blackbody emitter, providing high accuracy and repeatability. The temperature controller uses the industry standard PID algorithms to control the emitter temperature to within 0.1℃ of the set points. The black body emitter mechanism uses a resistive heating device that provides a long life, short stabilization times, and stable temperature control.
SPECIFICATION
- Temperature Range: 30℃ to 50℃
- Aperture size: 72 mm * 72mm
- Emissivity: 0.95
- Display resolution: 0.1℃
- Control stability:0.1℃ after 10 minutes
- Blackbody uniformity: 0.2℃
- Heat cavity shape:Flate Plate
- Warm up time:2 -10 minutes from ambient to 37℃
- Control method: PID auto control
- Temperature sensor: Pt100 class A
- Power Supply: 24VAC, 90W
- Operating ambient temperature: 0℃ to 40℃
- Operating humidity: < 90% (no condensation)
- Dimension w/o tripod:110mm (W) x 110mm(H) x 180mm(D)
- Net Weight: 2 kg
REFERENCE
- IEC TS 62492 – 2 : 2013 Industrial process control devices – Radiation thermometers – Part 2: Determination of the technical data for radiation thermometers”
- Standard ASTM E1311-14 “Standard Test Method for Minimum Detectable Temperature Difference for Thermal Imaging Systems”
- Standard ASTM E1213-14 “Standard Test Method for Minimum Resolvable Temperature Difference for Thermal Imaging Systems”
- Standard ASTM E1543-14 “Standard Test Method for Noise Equivalent Temperature Difference of Thermal Imaging Systems”
- International recommendation OIML R 141, 2008“Procedure for calibration and verification of the main characteristics of thermographic instruments”
- IEC TS 63144-1“Industrial process control devices -Thermographic cameras – Part 1: Metrological characterization”