Specifications
Specifications | |||
Maximum voltage between any terminal and earth ground | 600 V | ||
Surge protection | 6 kV peak per IEC 61010–1 600 VCAT III, Pollution Degree 2 | ||
Display | Digital: 6,000 counts, updates 4 per second | ||
Bar graph | 33 segments, updates 32 per second | ||
Operating temperature | -10°C to + 50°C | ||
Storage temperature | -40°C to + 60°C | ||
Battery type | 9 volt Alkaline, NEDA 1604A/ IEC 6LR61 | ||
Battery life | 400 hours typical, without backlight | ||
Accuracy Specifications | |||
DC millivolts | Range/resolution | 600.0 mV / 0.1 mV | |
Accuracy | ± ([% of reading] + [counts]): 0.5% + 2 | ||
DC volts | Range/resolution | 6.000 V / 0.001 V 60.00 V / 0.01 V 600.00 V / 0.1 V | |
Accuracy | ± ([% of reading] + [counts]): 0.5% + 2 | ||
AC millivolts1 True RMS | Range/resolution | 600.0 mV / 0.1 mV | |
Accuracy | 1.0% + 3 (DC, 45 Hz to 500 Hz) 2.0% + 3 (500 Hz to 1 kHz) | ||
AC volts1 True RMS | Range/resolution | 6.000 V / 0.001 V 60.00 V / 0.01 V 600.0 V / 0.1 V | |
Accuracy | 1.0% + 3 (DC, 45 Hz to 500 Hz) 2.0% + 3 (500 Hz to 1 kHz) | ||
Continuity | Range/resolution | 600 Ω / 1 Ω | |
Accuracy | Beeper on < 20 Ω, off > 250 Ω detects opens or shorts of 500 μs or longer. | ||
Ohms | Range/resolution | 600.0 Ω / 0.1 Ω 6.000 kΩ / 0.001 kΩ 60.00 kΩ / 0.01 kΩ 600.0 kΩ / 0.1 kΩ 6.000 MΩ / 0.001 MΩ | |
Accuracy | 0.9% + 1 | ||
Range/resolution | 40.00 MΩ / 0.01 MΩ | ||
Accuracy | 5% + 2 | ||
Diode test | Range/resolution | 2.00 V / 0.001 V | |
Accuracy | 0.9% + 2 | ||
Capacitance | Range/resolution | 1000 nF / 1 nF 10.00 μF / 0.01 μF 100.0 μF / 0.1 μF 9999 μF / 1 μF 100 μF to 1000 μF | |
Accuracy | 1.9% + 2 | ||
Range/resolution | > 1000 μF | ||
Accuracy | 5% + 20% | ||
Lo-Z capacitance | Range | 1 nF to 500 μF | |
Accuracy | 10% + 2 typical | ||
AC amps True RMS (45 Hz to 500Hz) | Range/resolution | 6.000 A / 0.001 A 10.00 A / 0.01 A | |
Accuracy | 1.5% + 3 20 A overload for 30-seconds max | ||
DC amps | Range/resolution | 6.000 A / 0.001 A 10.00 A / 0.01 A | |
Accuracy | 1.0% + 3 20 A overload for 30-seconds max | ||
Hz (V or A input)2 | Range/resolution | 99.99 Hz / 0.01 Hz 999.9 Hz / 0.01 Hz 9.999 kHz / 0.001 kHz 50 kHz / 0.01 kHz | |
Accuracy | 0.1% + 2 | ||
1. All AC voltage ranges except Auto-V/LoZ are specified from 1% to 100% of range. Auto-V/LoZ is specified from 0.0 V. 2. Frequency is AC coupled, 5Hz to 30 kHz for AC voltage. 2. Temperature uncertainty (accuracy) does not include the error of the thermocouple probe. 2. Frequency is AC coupled, 5 Hz to 50 kHz for AC voltage. Frequency is DC coupled, 45 Hz to 5 kHz for AC current. 2. Temperature Range/Resolution: -40°F to 752°F / 0.2°F. | |||
Mechanical & General Specifications | |||
Size | 167 x 84 x 46 mm (6.57 x 3.31 x 1.82 in) | ||
Weight | 550 g |